An Estimation of X-Radiation Output using Mathematic Model
- 1 Department of Radiologic Technology, Faculty of Associated Medical Sciences, Chiang Mai University, Chiang Mai 50200, Thailand
Abstract
Problem statement: Diagnosis x-ray radiation safety is key in medical examination. The quantity of patient radiation doses is beneficial for radiation protection of the patient. It was proposed that the equation for estimating the output (milliReongent, mR) from x-ray machines. Approach: A was 0.5, 0.8 and 1.0 for single phase, three phases and high frequency x-ray machines, respectively. To compare calculated output (mR) used this equation and measured output (mR) used ionizing chamber dosimeter. Results: The difference between the calculated and measured radiation dose was quite small. Conclusion: This equation could use to estimate output and it altered the reliable and inexpensive techniques for patient dose measurement in routine diagnostic x-ray examinations.
DOI: https://doi.org/10.3844/ajassp.2011.923.926
Copyright: © 2011 Suchart Kothan and Montree Tungjai. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
- 3,957 Views
- 10,446 Downloads
- 21 Citations
Download
Keywords
- Radiation output
- Ionization Chamber (IC)
- x-ray machine
- radiation dose
- calculated radiation